Bionet Mainstream EtCO₂ Sensor
EtCO2 Monitoring Module
The Bionet Mainstream EtCO₂ Sensor provides continuous measurement of end-tidal CO₂, inspired CO₂, and respiration rate directly at the patient’s airway. Its fast response and compact, lightweight design make it suitable for monitoring veterinary patients during anesthesia, surgery, recovery, and critical care.
The sensor uses dual-wavelength, non-dispersive infrared technology with no moving parts. A capnogram is displayed in less than four seconds, allowing clinicians to quickly assess ventilation and confirm airway status. The heated probe helps reduce condensation inside the airway adapter for more consistent monitoring.
The Bionet Mainstream EtCO₂ Sensor is compatible with Bionet Brio XVet Series and BM Elite Series patient monitors.
The Bionet Mainstream EtCO₂ Sensor provides continuous measurement of end-tidal CO₂, inspired CO₂, and respiration rate directly at the patient’s airway. Its fast response and compact, lightweight design make it suitable for monitoring veterinary patients during anesthesia, surgery, recovery, and critical care.
The sensor uses dual-wavelength, non-dispersive infrared technology with no moving parts. A capnogram is displayed in less than four seconds, allowing clinicians to quickly assess ventilation and confirm airway status. The heated probe helps reduce condensation inside the airway adapter for more consistent monitoring.
The Bionet Mainstream EtCO₂ Sensor is compatible with Bionet Brio XVet Series and BM Elite Series patient monitors.
Product Highlights
Fast
Capnogram displayed in less than 4 seconds, with a response time of less than 500 ms
Direct
Measures EtCO2, inspired CO2, and respiration rate directly at the patient’s airway
Accurate
Dual-wavelength infrared technology provides precise, continuous CO₂ measurement, with anesthetic agent compensation
Condensation Control
Heated probe helps reduce condensation inside the airway adapter for consistent monitoring
Lightweight
Compact, lightweight sensor design minimizes added weight at the patient’s airway
Low Maintenance
No routine gas calibration required. Backed by a 2-year warranty
Quick Link
Product Highlights
Fast
Capnogram displayed in less than 4 seconds, with a response time of less than 500 ms
Direct
Measures EtCO2, inspired CO2, and respiration rate directly at the patient’s airway
Accurate
Dual-wavelength infrared technology provides precise, continuous CO₂ measurement, with anesthetic agent compensation
Condensation Control
Heated probe helps reduce condensation inside the airway adapter for consistent monitoring
Lightweight
Compact, lightweight sensor design minimizes added weight at the patient’s airway
Low Maintenance
No routine gas calibration required. Backed by a 2-year warranty
Quick Link
The Bionet Mainstream EtCO₂ Sensor provides continuous measurement of end-tidal CO₂, inspired CO₂, and respiration rate directly at the patient’s airway. Its fast response and compact, lightweight design make it suitable for monitoring veterinary patients during anesthesia, surgery, recovery, and critical care.
The sensor uses dual-wavelength, non-dispersive infrared technology with no moving parts. A capnogram is displayed in less than four seconds, allowing clinicians to quickly assess ventilation and confirm airway status. The heated probe helps reduce condensation inside the airway adapter for more consistent monitoring.
The Bionet Mainstream EtCO₂ Sensor is compatible with Bionet Brio XVet Series and BM Elite Series patient monitors.
Intended Use: To monitor vital signs in veterinary patients (feline, canine), including: EtCO2
Key Features
- Mainstream CO₂ measurement directly at the airway
- Continuous EtCO₂, inspired CO₂, respiration-rate, and capnogram monitoring
- Capnogram displayed in less than four seconds
- 100 Hz sampling rate
- Anesthetic agent compensation (selectable)
- Respiration-rate range of 3 to 150 breaths per minute
- Compatible with adult and neonatal airway adapters
- Heated probe helps prevent condensation
- Lightweight sensor with a 102-inch cable
- Two-year warranty
- IP33-rated protection
Setup and Zero Calibration
The Bionet Mainstream EtCO₂ Sensor does not require routine gas calibration. However, an adapter-zero procedure is recommended before patient use.
For the most reliable results, connect the sensor and allow it to warm up for five minutes before performing the zero-calibration procedure. Complete the procedure with the appropriate airway adapter installed and exposed to room air. Do not perform zero calibration while the adapter is connected to the patient or breathing circuit.
General Specifications
| wdt_ID | Item | Specification |
|---|---|---|
| 73 | Bionet part number | B-RMCO2-C |
| 74 | Sensor type | Mainstream CO₂ |
| 75 | Compatible monitors | Brio XVet Series and BM Elite Series |
| 77 | Measurement principle | Dual-wavelength non-dispersive infrared, with no moving parts |
| 78 | Initialization time | Capnogram displayed in less than 4 seconds |
| 79 | Sampling rate | 100 Hz |
| 80 | Anesthetic agent compensation | Manually selectable |
| 82 | Weight | 0.18 lbs |
| 83 | Cable length | 102 inches |
| 84 | Warranty | 2 years |
| 85 | IP rating | IP33 |
| 86 | Routine gas calibration | Not required |
| 87 | Adapter-zero procedure | Required |
| 88 | Recommended warmup before zeroing | 5 minutes |
CO₂ Measurement Performance
| wdt_ID | Item | Specification |
|---|---|---|
| 73 | Measurement range | 0–114 mmHg, 0–15% CO₂, or 0–15.2 kPa |
| 74 | Rise time, T10–T90 | Less than 120 ms with adult or neonatal reusable or disposable airway adapters |
| 75 | Response time | Less than 500 ms |
| 77 | Resolution from 0–59 mmHg | 0.2 mmHg |
| 78 | Resolution from 60–114 mmHg | 0.5 mmHg |
| 79 | Noise | RMS ≤0.5 mmHg at 5% CO₂ |
CO₂ Accuracy
| wdt_ID | CO₂ range | Accuracy |
|---|---|---|
| 73 | 0–40 mmHg | ±2 mmHg |
| 74 | 41–70 mmHg | ±5% of reading |
| 75 | 71–100 mmHg | ±8% of reading |
| 77 | 101–114 mmHg | ±10% of reading |
| 78 | Respiration rates above 80 breaths/min | ±12% of reading |
Accuracy specifications are based on a gas temperature of 25°C.
Respiration Monitoring
| wdt_ID | Item | Specification |
|---|---|---|
| 73 | Respiration-rate range | 3–150 breaths/min |
| 74 | Respiration-rate accuracy | ±1 breath/min |
| 75 | EtCO₂ calculation | Peak of the expired CO₂ waveform |
| 77 | Inspired CO₂ calculation | Lowest CO₂ value measured during the previous 20 seconds |
Stability
| wdt_ID | Item | Specification |
|---|---|---|
| 73 | Short-term drift | ≤1 mmHg over 3 hours |
| 74 | Long-term drift | Accuracy maintained over 6 hours |
Probe Heating
| wdt_ID | Item | Specification |
|---|---|---|
| 73 | Purpose | Helps prevent condensation inside the airway adapter |
| 74 | Target temperature | 37°C |
| 75 | Maximum heating power | 0.5 W |
| 76 | Temperature protection | Heating automatically stops if the internal temperature exceeds 37°C |
* Specifications can be changed without prior notification.
- Overview
-
The Bionet Mainstream EtCO₂ Sensor provides continuous measurement of end-tidal CO₂, inspired CO₂, and respiration rate directly at the patient’s airway. Its fast response and compact, lightweight design make it suitable for monitoring veterinary patients during anesthesia, surgery, recovery, and critical care.
The sensor uses dual-wavelength, non-dispersive infrared technology with no moving parts. A capnogram is displayed in less than four seconds, allowing clinicians to quickly assess ventilation and confirm airway status. The heated probe helps reduce condensation inside the airway adapter for more consistent monitoring.
The Bionet Mainstream EtCO₂ Sensor is compatible with Bionet Brio XVet Series and BM Elite Series patient monitors.
Intended Use: To monitor vital signs in veterinary patients (feline, canine), including: EtCO2
Key Features
- Mainstream CO₂ measurement directly at the airway
- Continuous EtCO₂, inspired CO₂, respiration-rate, and capnogram monitoring
- Capnogram displayed in less than four seconds
- 100 Hz sampling rate
- Anesthetic agent compensation (selectable)
- Respiration-rate range of 3 to 150 breaths per minute
- Compatible with adult and neonatal airway adapters
- Heated probe helps prevent condensation
- Lightweight sensor with a 102-inch cable
- Two-year warranty
- IP33-rated protection
Setup and Zero Calibration
The Bionet Mainstream EtCO₂ Sensor does not require routine gas calibration. However, an adapter-zero procedure is recommended before patient use.
For the most reliable results, connect the sensor and allow it to warm up for five minutes before performing the zero-calibration procedure. Complete the procedure with the appropriate airway adapter installed and exposed to room air. Do not perform zero calibration while the adapter is connected to the patient or breathing circuit.
- Specifications
-
General Specifications
wdt_ID Item Specification 73 Bionet part number B-RMCO2-C 74 Sensor type Mainstream CO₂ 75 Compatible monitors Brio XVet Series and BM Elite Series 77 Measurement principle Dual-wavelength non-dispersive infrared, with no moving parts 78 Initialization time Capnogram displayed in less than 4 seconds 79 Sampling rate 100 Hz 80 Anesthetic agent compensation Manually selectable 82 Weight 0.18 lbs 83 Cable length 102 inches 84 Warranty 2 years 85 IP rating IP33 86 Routine gas calibration Not required 87 Adapter-zero procedure Required 88 Recommended warmup before zeroing 5 minutes CO₂ Measurement Performance
wdt_ID Item Specification 73 Measurement range 0–114 mmHg, 0–15% CO₂, or 0–15.2 kPa 74 Rise time, T10–T90 Less than 120 ms with adult or neonatal reusable or disposable airway adapters 75 Response time Less than 500 ms 77 Resolution from 0–59 mmHg 0.2 mmHg 78 Resolution from 60–114 mmHg 0.5 mmHg 79 Noise RMS ≤0.5 mmHg at 5% CO₂ CO₂ Accuracy
wdt_ID CO₂ range Accuracy 73 0–40 mmHg ±2 mmHg 74 41–70 mmHg ±5% of reading 75 71–100 mmHg ±8% of reading 77 101–114 mmHg ±10% of reading 78 Respiration rates above 80 breaths/min ±12% of reading Accuracy specifications are based on a gas temperature of 25°C.
Respiration Monitoring
wdt_ID Item Specification 73 Respiration-rate range 3–150 breaths/min 74 Respiration-rate accuracy ±1 breath/min 75 EtCO₂ calculation Peak of the expired CO₂ waveform 77 Inspired CO₂ calculation Lowest CO₂ value measured during the previous 20 seconds Stability
wdt_ID Item Specification 73 Short-term drift ≤1 mmHg over 3 hours 74 Long-term drift Accuracy maintained over 6 hours Probe Heating
wdt_ID Item Specification 73 Purpose Helps prevent condensation inside the airway adapter 74 Target temperature 37°C 75 Maximum heating power 0.5 W 76 Temperature protection Heating automatically stops if the internal temperature exceeds 37°C * Specifications can be changed without prior notification.